  • iCount 3CP
  • ICount3 TS
  • iCount 5

Features

3-part WBC Diff Hematology analyser

  • 22 parameters including WBC, LYM, MID, GRA, LYM%, MID%, GRA%, RBC, MCV, HCT, HGB, MCH, MCHC, PLT, MPV, RDW-SD, RDW-CV, PCT,PDW-SD, PDW-CV, P-LCR, P-LCC
  • Open and closed tube sampling with automatic sample rotor and Cap Piercing
  • Throughput of 60 tests / hour
  • 3 separate chamber for primary mixing, RBC and WBC counting
  • High quality cyanide free reagents imported from Europe
  • 6 QC Level, 16 and 64-day Levey-Jennings charts, separate QC database
  • Data storage capacity of 10000 results with RBC, PLT, WBC 3-part histogram
  • Optional external barcode reader and keyboard
  • High-contrast, backlit, color graphics LCD
  • Capability of using any external printer via a printer box
  • US-FDA approved

Hematology diagnosis is not a matter of compromise. iCount3CP combines reliability and accuracy with computer-enhanced benefits of handling. iCount3CP is a heavy duty compact analyzer offering an optimal solution for hospitals, clinics and practices, which require high quality and fast CBC and 3-part WBC differential results. Our Hematology analyzers bring the convenience and accuracy of the reference laboratory right into your practice, offering maximum value at low running costs.

Hematology Analyser
Hematology Analyzer
3-part WBC Diff Hematology Analyser

High Performance

  • iCount3CP provides 20-parameters, a complete CBC including a 3 part WBC differential
  • High throughput of 60 samples per hour, results within 60 seconds
  • The fluidic system is automatically cleaned from dust particles when turning on the analyzer. High voltage pulses after each measuring cycle keep the apertures free from clogging. The instrument automatically cleans the sampling tip inside and outside after blood aspiration to minimize the risk of carryover and infection.
Hematology Analyser Performance
Hematology Analyzer Performance

Safe

  • Cap piercing feature to enable accurate sampling from closed tubes thus minimising biohazard risks.

Ease of use

  • Color LCD, the large display allows rapid and accurate patient data entry and monitoring of the results with large histograms
  • Sample rotor, ensuring safe and easy sampling
  • The instrument features a sophisticated but easy-to-use, logically constructed multilingual operating menu. It provides 6-level QC, L-J graphs and self-diagnostic functions to let you monitor reliability and accuracy
  • All iCount3CP models can be interfaced with your host computer for further processing of the measured data. Results can also be retrieved on the built-in roll paper printer

Reliable and efficient

  • iCount3CP is almost maintenance free resulting in minimal disruption of operation
  • iCount analyzers are small and light. Their compact size makes the instruments ideal for use in smaller clinics requiring minimum desktop space
  • Clinical environment by utilizing industry standard electronic and mechanic components
Reliable Hematology Analyzer
Reliable Hematology Analyser
Efficient Hematology Analyzer

Technical Specifications

Sample method Closed and open tube sampling with automatic sample rotor
Parameters 20 parameters, including 3-part WBC differential: WBC,LYM#, MID#, GRA#, LYM%, MID%, GRA% RBC, MCV, HCT, HGB, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PCT, PDW-CV, PDW-SD
Chambers 3 chambers for diluting whole blood and counting; 1:MIX, 1:RBC, 1:WBC
Reagent system Isotonic Diluent, Lyse, Cleaner
Aperture diameter 80 ìm (RBC/PLT), 100 ìm (WBC)
Throughput 60 tests/hour
Sample types Human(general), male, female, baby, toddler and child (built-in reference ranges)
Clog prevention High-voltage pulse on aperture in each analysis cycle; chemical cleaning; high pressure backflush of aperture with Cleaner reagent.
Cleaning procedure High-voltage burst of the aperture, high-pressure back-flush, chemical cleaning of the aperture using Cleaner reagent.
Quality control 6 QC levels, QC parameters include: mean, ± range, SD and CV for all measured and calculated parameters, 16- and 64-day Levey-Jennings charts, separate QC database.
Calibration 1 or 3-measurement automatic and manual (factors) calibration of WBC, HGB, RBC, PLT, MCV, RDW and MPV absolute. Independent calibration of pre-diluted mode.
Multi-user feature (advanced feature) 3-level multi-user operation with selective privilege levels, user identification with ID and password.
User interface Easy-to-use, menu driven user interface with 6 software buttons (with graphic icons), 6 hardware function buttons, cursor and numeric keys.
Languages available English, German, Hungarian, Arabic, Bulgarian, French, Greek, Indonesian, Polish, Portuguese, Romanian, Russian, Slovakian, Spanish, Turkish, Vietnamese
Data capacity Maximum 5000 results with RBC,PLT, WBC 3-part histogram
Host computer interface USB B port or RS-232 serial link.
Data back-up method USB mass storage device; Lab data management software.
Software upgrade method USB A port using USB mass storage device
Printer interface USB with support for HP printers, color and monochrome printout.
Built-in printer "Easy Paper Operation" thermal printer module, full report with histograms
Display 320x240 -dots, high-contrast, backlit, color graphics LCD
Keypad 29 foil keys + separate START button, red/green/orange machine state light.
External keyboard Standard PS/2 compatible or USB
Power requirement 12VDC, 6A, 72W max. operating power.
Power supply unit External, auto-ranging power unit for 100-120 or 200-240 VAC, 50-60Hz.
Operating temperature 59-86 °F (15-30 °C). Optimal temperature is 77 °F (25 °C).
Dimensions (W x D x H) 350 X 385 X490 mm
Net weight 15 kg.

Accurate & Reliable Results

  • 21 parameters and RBC, WBC, PLT histograms
  • Utilize digital filtering, blood cell - overlap calibration, fitting curve, floating threshold technology etc. to enhance the accuracy and reliability of results
  • High quality syringe pump with high quantitative precision

Perfect QC Function

  • L-J, X, X-B 3 levels of controls for daily QC with Levey-Jennings charts

Easy Maintenance

  • Zap and backflush cleaning system to prevent apertures blocking
  • Special boot program to rinse pipelines and probe automatically
  • Only 3 reagents involving test

Large Storage

  • 10,000 results with 3 histograms to satisfy customers' requirements

User Friendly Design

  • Cyanide-free hemolytic reagent for HGB detection, safety and environment pollution free
  • Eiectric hydraulic separation structure design
  • Automatic reagents remaining detection
  • USB and RS232 interfaces, easy for backup
  • Multiple languages
  • LIS connection

Easy Operation

  • Large color LCD with high resolution.
  • Automatic sampling, mixing and analyzing
  • Equipped with barcode reader
  • Abundant prompt information
  • Abnormal value flag and reference ranges modified as the local requirements
  • Automatic dormancy setting
Haematology ICount3 TS

Technical Specifications

Parameters WBC, RBC, HGB, MCV, MCH, MCHC, HCT,PLT, NEUT#, MXD#, LYM#, NEUT%,MXD%,LYM%,R DW-CV, ROW-SO, POW, MPV,PCT, P-LCR, P-LCC
Histogram WBC, RBC, PLT
Principles WBC, RBC, PLT : electrical impedance HGB: Cyanide-free colorimetry
Cross Contamination WBC, RBC, HGB < 0.5%, PLT < 1%
Aperture Diameter WBC 100 μm, RBC/PLT 70 μm
Sample Consumption Whole blood 6μL,Pre-diluted blood 20μL
Throughput 60 samples/hour
Display 10.1 inch touch screen, Resolution: 1024 X 600
Reagent Detergent, Diluents, BCC-3600 Lyse
Multiply QC methods L-J, X:12QC files,30 results/file.X-B:1file, 150 results/file
Calibration Manual calibration, automatic calibration,fresh blood calibration
Counting Mode Whole blood and pre-diluted blood
Input/ Output Build in thermal printer,RS232,USB
Environment Requirement Ambient temperature 18°C-30°C,Relative humidity: ≤ 75%
Power Supply 100-240 VAC 50/60HZ
Dimension (mm) 430(L) X 395(W) X 432(H)
Weight 25 Kg

iCount5 is one of the high-end models of iCount Hematology Analyzers range. The ‘iCount5’ offers an optimal solution for hospitals, clinics and practices, which require a high throughput 24 parameter analyzer with laser based optical measuring technology for precise and accurate 5-part differential results for human samples. Our Hematology analyzers bring the convenience and accuracy of the reference laboratory right into your practice, offering maximum value at low running costs.

Haematology iCount 5
Haematology Analyzers iCount 5
Haematology Analyser iCount 5

High performance

  • Complete 24-parameter CBC profile including the optical determination of the 5 part WBC differential count
  • 60 samples per hour throughput (single runs or work order)
  • Auto-sampler with integrated bar code reader available
  • Diagnostic flagging system
  • Intelligent software is used for determining the different white blood cell populations

Ease of use

  • Color touch-screen, allowing fast and efficient work and monitoring of the results with large histograms and scattergrams
  • Sophisticated operating menu, the instrument features an easy-to-use,logically constructed multilingual operating menu. It provides 6-level QC, L-J graphs and self-diagnostic functions to let you monitor reliability and accuracy
  • Integrated PC enabling easy connection to any (LIS) system or printer
  • Reporting on any Windows compatible printer

Reliable and efficient

  • The optical measurement technology utilizes laser diode ensuring a long lifetime and precise results with the special, high-tech flow cuvette
  • iCount5 requires minimal maintenance, resulting in minimal disruption of operation
  • iCount analyzers are kept small and light. Their compact size makes the instruments ideal for use in small to medium sized clinics requiring minimum desktop space
  • Precise and reliable 5-part hematology profiles are guaranteed even in the most demanding clinical environment by utilizing industry standard electronic and mechanic components

Technical Specifications

Sample volume Closed-, and Open-mode: 100 μl
Sample type Human whole blood (K-EDTA anticoagulant)
Tube Identification By means of the front panel keyboard (enter ID) By means of the barcode labels (manual and/or auto-sampler)
Sampling method Ceramic shear valve with 3 separated primary loops
Measured parameters CBC+5 DIFF mode ( 24 parameters): WBC, LYM, MON, NEU, EOS, BAS, LYM%, MON%, NEU%, EOS%, BAS%, RBC, HCT, MCV, RDW-CV, RDW-SD, HGB,MCH, MCHC, PLT, PCT, MPV, PDW-CV, PDW-SD CBC mode (12 parameters): WBC, RBC, HCT, MCV, RDW, HGB, MCH, MCHC, PLT, PCT, MPV, PDW
Throughput 60 tests/hour
Measurement method
  • Volumetric impedance change for WBC, RBC, PLT. Spectrophotometry for HGB
  • Light scattering 4-diff measurement: LYM, MON, NEU, EOS
  • Light scattering BASO measurement
Aperture diameter WBC: 100 μm, RBC/PLT: 80 μm
Aperture length WBC: 100 ìm, RBC/PLT: 80 ìm
HGB measurement Light source : green LED with 540 nm wavelength. Detector: light to frequency converter
Optical measurement
  • Light source : semiconductor laser diode with 650 nm wavelength and 10mW
  • (Class IIIB laser module if the protective housing is closed)
  • Quartz flow cell with hydro-dynamic focusing
  • Detector: fiber optic coupled PIN Si photodiodes Internal safety interlock
Auto-alignment system
  • Optional. Horizontal and vertical calibration of laser beam path
  • Coarse calibration: with blood
  • Fine calibration: with calibration material (Polystyrene micro particle or Polystyrene microsphere, 7μm)
Reagents
  • Diatro • Dil-DIFF (20 liter)
  • Diatro • Lyse-5P (5 liter)
  • Diatro • Diff-5P (5 liter)
  • Diatro • Hypocleaner CC (100 ml) (Emergency cleaner)
Dilution ratios
  • WBC/BAS 1: 228
  • RBC/PLT 1: 32.000
  • 4 DIFF 1: 250
Sheath fluid Diluent
Control material D-Check 3P, Manufacturer: R&D Systems
Quality Control 16- and 64-day Levey-Jennings charts, separate QC database (6 level).
Flagging Pathological (diagnostic) flags. Lab limits (normal ranges). Reagents alert (3 measurement prealert-online reagent replacement) Instrument alerts, internal puffer for reagents
Calibration Manual and SW supported automatic mode
Languages available English menu and support for other languages
Software upgrade Via USB
Data storage capacity 100.000 records including flags, scatter- and histograms
Data processing VIA C7 1.8 GHz processor
Data store Embedded XP
Display 800 x 600 color graphic LCD, portrait layout
External printing Via USB port, any Windows compatible printer
External keyboard Via PS/2 or USB
Bar-Code reader Optional Manual bar-code reader via USB Built in Bar-Code in the Auto-samplerB
Peripheral ports USB (2.0) 4pc., Ethernet, PS/2
Power requirements
  • Power supply input: from 90-135Vac to 180-265Vac; 47Hz to 63Hz
  • Power supply input current: <10A @ 115Vac; <5A @ 230Vac
  • 0 0 Power Consumption: maximum 350 VA and15-35 C (59-98 F);
Operating temperature Maximum relative humidity 80%